Minimum detectable activity (MDA) is the lowest activity of a radioactive nuclide that can be detected with a detector, to some confidence level.[1][page needed] It is a concept that is used in several circumstances, such as in whole-body counting or radiation monitoring, to aid in determining the presence or absence of a radioactive substance or comparing the performance of different detector systems. There are several ways to calculate MDA that are commonly used, including the ISO11929 standard and the Currie method.[2] The Currie method is given by[3]

,

where is the detection limit in units of counts, is the detection efficiency of the detector, is the emission probability of the radiation, and is the live time of the measurement. The formula for will vary depending on the specified confidence level. For false-negative and false-positive rates of 5%,[1]

,

where is the background counts. In other words, to achieve a low MDA, low background counts and a high efficiency are desired. This is usually referred to as the Currie detection limit.

Methods to decrease MDAs include utilizing coincidence measurements,[4] large detection volumes,[5] or using large shielding and underground measurement facilities.[6]

Currie detection limit - derivation

The Currie detection limit is derived from the number of gross counts G and the number of background counts B and an accepted false positive rate of 5%[7]. The derivation assumes that both G and B follow a Gaussian distribution.

The number of net counts N in a peak is given as

Using that G and B follow a Gaussian distribution, the standard deviation of N is

If the number of counts N is below a critical limit , the conclusion is that no detection is made. The value of depends on the background and is derived by observing the case where only background activity is present. In this case, the mean of N is zero and an uncertainty is

since if there is only background.

To achieve a false positive rate of 5%, must be chosen such that the probability that a random sample lies below is 95%. Since B follows a Gaussian distribution, it is known that 90% of values lie within 1.64 sigma of the mean. In this case, only positive deviation are of concern and the probability is 95% that a random sample lies below the mean plus . Thus, the critical limit can be set to

Looking now at the case where there is activity present that does not belong to the background, the detection limit can be determined. The detection limit is determined so that 95% of samples from the net count distribution N lies above the critical limit

Assuming N follows a Gaussian distribution where the width is only caused by statistical uncertainties from counting, setting N to (minimum amount of counts that will lead to a detection) gives

Using this together with the definition of results in

Which after solving for gives

References

  1. ^ a b Gilmore, Gordon (2011). Practical gamma-ray spectrometry (2., repr. with corr ed.). Chichester: Wiley. ISBN 978-0-470-86196-7.
  2. ^ Done, L.; Ioan, M-R. (2016). "Minimum Detectable Activity in gamma spectrometry and its use in low level activity measurements". Applied Radiation and Isotopes. 114: 28–32. Bibcode:2016AppRI.114...28D. doi:10.1016/j.apradiso.2016.05.004. ISSN 0969-8043. PMID 27172893.
  3. ^ Currie, L. A. (1968). "Limits for qualitative detection and quantitative determination: Application to radiochemistry". Analytical Chemistry. 40 (3): 586–593. Bibcode:1968AnaCh..40..586C. doi:10.1021/ac60259a007. ISSN 0003-2700.
  4. ^ Britton, R.; Davies, A. V.; Burnett, J. L.; Jackson, M. J. (2015). "A high-efficiency HPGe coincidence system for environmental analysis". Journal of Environmental Radioactivity. 146: 1–5. Bibcode:2015JEnvR.146....1B. doi:10.1016/j.jenvrad.2015.03.033. ISSN 0265-931X. PMID 25875083.
  5. ^ Keyser, R. M.; Twomey, T. R.; Wagner, S. E. (1990). "The Benefits of Using Super-Large Germanium Gamma- Ray Detectors for the Quantitative Determination of Environmental Radionuclides" (PDF). Ortec. Retrieved 10 November 2025.
  6. ^ Laubenstein, M.; Hult, M.; Gasparro, J.; Arnold, D.; Naumaier, S.; Heusser, G.; Köhler, M.; Povinec, P.; Reyss, J.-L.; Schwaiger, M.; Theodórsson, P. (2004). "Underground measurements of radioactivity". Applied Radiation and Isotopes. 61 (2–3): 167–172. Bibcode:2004AppRI..61..167L. doi:10.1016/j.apradiso.2004.03.039. ISSN 0969-8043. PMID 15177339.
  7. ^ Knoll, Glenn F. (2010). Radiation detection and measurement (4th ed.). Hoboken, N.J: John Wiley. pp. 96–98. ISBN 978-0-470-13148-0.